Portable Pipet-Aid®

Portable Pipet-Aid®

Regular price $409.00
Regular price Sale price $409.00
Sale Contact for Availablilty
Shipping calculated at checkout.

Drummond Scientific's Portable Pipet-Aid® is a must-have precision device for technicians and specialists worldwide. This reliable and durable tool ensures consistent and accurate results, making it an indispensable part of any laboratory. With cordless functionality and a flow rate of better than 25 mL in 5 seconds when using a 100 mL pipette, the Portable Pipet-Aid® is the go-to choice for professionals who demand the best.

View full details

Portable Pipet-Aid®

Additional Resources

Specifications

Size: 10″
Weight:
8 oz.
Capacity:
Accepts 1 to 100 mL pipets
Battery:
Ni-MH (3.6V/750 mAh)

Operations & Service

All portable units use a Ni-MH rechargeable battery. Recharge the unit when not in operation. Routine overnight charging has no detrimental effect on the battery. When fully charged, the unit will handle a full days’ pipetting requirements. Replacement batteries are supplied by Drummond and are matched to our charging units. Do not use another manufacturers’ charger. In normal use, the batteries should last at least two years.

If the battery charge is low, operating the unit with the charger plugged in will help, however, if the battery is dead, plugging in the charger will not help. Check apparent “dead” battery with another Drummond charger.

CHANGING THE BATTERY: (Watch a step-by-step video on the “Maintainance Video” tab) to replace battery.

(1) Remove entire nosepiece by unscrewing. (1A) Older models with “clip-on” nosepieces must have the T.C. bracket removed also. To do this, remove the T.C. gasket which usually remains in the upper portion to the handle. Take a coin and unscrew the T.C. screw to remove the bracket.

Once removed, (2) cut the label on the bottom of the handle near the charger terminal.

(3) Remove the 5 screws on the handle.

Once removed, (4) place the unit flat on a counter with the SCREW HOLES DOWN.

(5) Lift the top portion off and the internal components will remain in the bottom portion of the handle.

(6) The battery will be seen. Removal may require the gentle prying with a small screw driver to remove. It is held in with double-backed tape.

(7) Unplug battery from connector, replace with new Ni-MH Drummond battery (4-000-035) available from your local laboratory supply dealer or from Drummond directly.

If you experience any difficulty, call Drummond’s HOTLINE at: 1-800-523-7480

T.C. Nosepieces do not use any “cotton” filters. Remove cotton found in any nosepiece. The Drummond Self-locking filter found in all portable nosepieces (and some standard Pipet-Aids), provides a unique protection against overpipetting. One drop of aqueous liquid will cause this filter to block. Replacement is necessary. Other filters do not offer this protection. DO NOT USE SUBSTITUTES. Wetting the filter will prevent any liquid flow in the pipette; however, the pump will sound normal. Replacement of the filter will restore normal function. CAUTION: When replacing the filter, make sure the rubber insert is clean and dry. A “hang-up” drop of liquid in the rubber insert can ruin the new filter.

Warranty Form

Frequently Asked Questions

Why is there no aspiration or pressure when the buttons are pressed?

Original Pipet-Aid – 4-000-110/4-000-111

1. Diaphragm inside pump is split or dried up and cracked. Replace diaphragms (contact us)

2. Check valve in nosepiece needs to be cleaned out. Push wire or paper clip through black O-ring to open up check valve. When opened, rinse float and shell with water or alcohol. Allow to dry and reassemble. Shake and listen for float to rattle. Reassemble into nosepiece and check for aspiration.

Original Pipet-Aid – 4-000-110-TC/4-000-111-TC

1. Filter has been wetted. Replace with 4-000-051.

2. TC gasket missing in nosepiece, requires one on each end of filter

3. Diaphragm inside pump is split or dried up and cracked. Replace diaphragms.

Original Portable Pipet-Aid – 4-000-100/4-000-200

1. Filter has been wetted. Replace with 4-000-051.

2. TC gasket missing in nosepiece, requires one on each end of filter (4-000-057).

Portable Pipet-Aid XP/XP Gravity – 4-000-101/4-000-102, 4-000-201/4-000-202

1. Filter has been wetted. Replace with 4-000-051.

a. Pull back the nose-piece release and turn the nose piece counterclockwise.

2. TC gasket missing in nosepiece, requires one on each end of filter (4-000-057).

3. Fill or “dispense” button between speed settings. Rotate until you feel it click into place.

4. Fill or “dispense” button is broken. Check to make sure the buttons click into place; they should not spin freely.

5. Cotton plug in rubber insert blocking flow.

6. Connector tube inside the handle has come undone.

Hoodmate Pipet-Aid – 4-000-300, 4-000-302, 4-000-303

1. Filter has been wetted. Replace with 4-000-051.

a. Pull back the nose-piece release and turn the nose piece counterclockwise.

2. TC gasket missing in nosepiece, requires one on each end of filter (4-000-057).

3. “Fill” or “dispense” button between speed settings. Rotate until you feel it click into place.

4. “Fill” or “dispense” button is broken. Check to make sure the buttons click into place; they should not spin freely.

5. Cotton plug in rubber insert blocking flow.

Portable Pipet-Aid XL – 4-000-105, 4-000-205

1. Filter has been wetted. Replace with 4-000-051. 

a. Press the nose-piece release and turn the nose piece counterclockwise.

2. TC gasket missing in nosepiece, requires one on each end of filter (4-000-057).

3. “Fill” or “dispense” button between speed settings. Rotate until you feel it click into place.

4. “Fill” or “dispense” button is broken. Check to make sure the buttons click into place; they should not spin freely.

5. Cotton plug in rubber insert blocking flow.

Why is the pipette not holding volume (leaking)?

Original Pipet – 4-000-110-TC/4-000-111-TC

1. Diaphragm inside pump is split or dried up and cracked. Replace diaphragms.

2. Replace rubber insert.

Portable Pipet-Aid XP/XP Gravity – 4-000-101/4-000-102, 4-000-201/4-000-202

1. Replace rubber insert (4-000-002) if worn
2. Poor seal around filter; use Drummond filters (4-000-051) and be sure both black, rubber
gaskets are on either end of the filter.
3. Handle connector tubing has a pin hole in it. Cut out if long enough or replace (4-000-013).

Portable Pipet-Aid XP2 – 4-000-501

1. Replace worn rubber insert (4-000-002)

2. Valve body O-rings need to be replaced.

3. “Fill” or “dispense” button is broken. Check to make sure the buttons click into place; they should not spin freely.

 

Portable Pipet-Aid XL – 4-000-105, 4-000-205

1. Replace worn rubber insert (4-000-002).

2. Valve body O-rings need to be replaced.

3. Pin-hole in tube inside handle.

What is the function of the Dual Filtration?

For use in a bio-hazardous environment, a filtration unit can be installed on the Pipet-Aid pump which will filter both vacuum and pressure lines. Incorporating dual filters, the Pipet-Aid Biohazard Filter Kit has a retention efficiency of 99.9999+% for 0.1 micron particles and droplets.

*Normally factory installed, the Filter Adapter Kit can be attached to existing unfiltered Pipet-Aid pumps.

Why is there no aspiration or pressure when the buttons are pressed?

Original Portable Pipet-Aid

4-000-100/4-000-200

1. Filter has been wetted. Replace with 4-000-051
2. TC gasket missing in nosepiece, requires one on each end of filter (4-000-057)

*Normally factory installed, the Filter Adapter Kit can be attached to existing unfiltered Pipet-Aid pump.

Why is there no sound from motor when buttons pressed?

Original Portable Pipet-Aid – 4-000-100/4-000-200

1. Battery needs to be charged overnight.
2. Battery dead. Replace with (4-000-035).
a. Cut the label on the bottom along the seam of the handle. Remove the nose piece followed by the 5 screws. Flip the unit over so that the screw holes are now facing down, then you can open up the top half of the Pipet-Aid. The battery is held down by
double-sided tape.

3. Listen for slight click when buttons are first pushed. If not heard, micro switch is broken, replace wire harness (4-000-046).

Portable Pipet-Aid XP/XP Gravity – 4-000-101/4-000-102, 4-000-201/4-000-202

1. Battery needs to be charged overnight.

2. Battery is dead. Replace with 4-000-035.

a. Cut the label on the bottom along the seam if the handle. Remove the nose piece

followed by the middle screw. The battery door can now be removed.

3. Battery damaged by use of non-Drummond charger. Replace battery (4-000-035). Charger will have Cat# 4-000-030 printed on it and should be black.

4. Unit works intermittently or doesn’t work at all

a. Wire harness needs to be replaced 

Portable Pipet-Aid XP2 4-000-501

Unit needs to be recharged. Plug in power supply/charger and use unit while it charges, use only power supply/charger 4-000-065. If LED does not light, replace battery (4-000-036).

Portable Pipet-Aid XL – 4-000-105, 4-000-205
1. Plug in power supply/charger and use unit while it charges. Use only power supply/charger 4-000-065. Battery dead. Replace with (4-000-036).

2. If battery will not hold a charge replace with new battery, (4-000-036).

Why is aspiration/fill very slow (no leaking)?

1. Rubber insert has cotton plugs blocking flow. Clean out cotton plugs or replace rubber insert (4-000-002).
2. Buttons in valve body dirty. Clean or replace (4-000-021).

Why is the pipette not holding volume (unit leaks)?

1. Replace rubber insert (4-000-002) if worn
2. Poor seal around filter; use Drummond filters (4-000-051) and be sure both black, rubber
gaskets are on either end of the filter.
3. Handle connector tubing has a pin hole in it. Cut out if long enough or replace (4-000-013).

How long will it take to charge my device?

Original Portable Pipet-Aid: Charging the unit for up to 5 hours should completely charge the battery. It is recommended to keep the charger plugged in whenever not in use; this will ensure optimum battery life. All rechargeable batteries have a limited life span. Generally each battery will last approximately 1000 cycles (charges) and then must be replaced.

Pipet-Aid XP/XP2: It should take up to two and a half hours to fully charge the battery. You can still use the Pipet-Aid while it’s plugged into the charger; the charger also seconds as a power supply.

Pipet-Aid XL: It should take up to two and a half hours to fully charge the battery. You can still use the Pipet-Aid while it’s plugged into the charger. The charger also seconds as a power supply.

What is the difference between the XP (4-000-101) and the XP Gravity (4-000-102)?

The Pipet-Aid XP features three speeds on the buttons; (S)low, (M)edium and (F)ast. On the XP
Gravity, the fast speed on the dispense button is replaced with a (G)ravity drain. This feature
allows the user to dispense the liquid by gravity; allowing the liquid to simply fall out of the
pipette when the button is pressed…the motor will not run in this mode.

Can the Pipet-Aid be autoclaved?

The short answer is no.

Although the nose piece and its components (with the exception of the
TC Filter) can be autoclaved.

– Wiping the unit with a soft cloth or wipe with isopropyl alcohol is sufficient to clean the
unit. 

– Avoid the use of acetone and other organic solvents which will damage the handle.

– Always disconnect the charger from the unit when cleaning.

– Prevents both over-pipetting and cross-contamination.

– Each layer has a pore size of .8µm; the two layers combined create an optimum barrier to particles as small as .1µm

– Once wetted, it will cease to pass air alerting the user that over pipetting has occurred.

What size pipettes can be used with the Pipet-Aid

Our unique rubber insert allows the user to use any serological, volumetric and/or blow-out pipettes in either glass, plastic or something similar with sizes ranging from 1 to 100mL.